특 징 |
|
High Sensitivity ; VUV-24 검출기는 0.0005~4.0 AU/V 감도범위의 고해상도 아나로그 출력신호를 가지며, ±0.00001 이하의 매우 낮은 노이즈 레벨을 나타냅니다. |
|
Wavelength Selection ; 중수소(deuterium, D2) 램프는 190nm ~ 400 nm의 파장범위의 광원을 제공하며, 키패드 또는 외부제어를 통해 단색화장치(prism)에 의한 특정파장을 선택할 수 있습니다.
|
|
Reliable, Rugged, Low Maintenance Design ; VUV-24 검출기는 신뢰성이 높은 중수소 램프와 진정한 이중빔 광원 및 견고한 전원공급장치로 구성되어 있으며, 기기의 피드백 멤브레인 키패드에 의해 제어됩니다. | | |
규 격 |
|
Light source |
|
Deuterium lamp, 190 nm - 400 nm, >2000 hour service life |
|
Optical design |
|
Double beam (sample/reference) |
|
Wavelength selection |
|
User selectable (front panel or RS232) in 1 nm increments |
|
Wavelength accuracy |
|
Better than ± 2 nm range |
|
Wavelength precision |
|
Better than ± 0.2 nm |
|
Band pass |
|
<8.0 nm (typical) at half-height @ 250 nm |
|
Noise* |
|
<± 1 x 10-5 AU over 30 sec period peak to peak with 1.0 sec rise time |
|
Drift* |
|
<± 1 x 10-4 AU over 30 min period peak to peak 1.0 sec rise time |
|
Input/Output |
|
RS-232 computer I/O; auto-injection (event or run); auto-zero input; one analog integrator or chart recorder output, (voltage range default 1.0 V/AU); selectable analog range outputs of 4.0, 2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.01, 0.005, 0.002, 0.001, 0.0005 AU/V with voltage range of ± 4V at any scale |
|
Display |
|
Absorbance (5 digit LED); wavelength (3 digit LED); analog output range (momentarily from absorbance display by pressing range switch); reference and sample light levels |
|
Size |
|
6.6" x 9.4" x 15" (HxWxD), Approx. 12 lb (shipping weight) | | |
주문정보 |
|
9024-0001 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, 10 mm STD cell, 18 microliter |
|
9024-0010 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, 5 mm STD cell , 10 microliter, low RI efffect (this cell availbale in low back pressure version) |
|
9024-0007 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, CAP cell , for 365 micron capillary with ID from 25 to 300 microns |
|
9024-0011 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, PREPFIX cell, 0 to 3 mm pathlength, set to 1.0 mm if not specified, very very low backpressure |
|
9024-0013 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, MICRO-1000 cell, 2 mm pathlength, 1 microliter volume |
|
9024-0009 |
Variable wavelength ultraviolet/visible absorbance monitor detector with D2 lamp 190 to 940nm, single wavelength monitoring, 10mm STD-PREP cell , low back pressure, high sensitivity |
|
Flow Cells |
|
|
9050-0001 |
STD flow cell for HPLC, 10 mm, 18 uL, 1000 psi |
|
9050-0010 |
STD flow cell for HPLC, 5 mm, 10 uL, 1000 psi |
|
9050-0007 |
CAP flow cell for micro-HPLC or CE, >50 um ID x 375 um OD capillary, "on-column" |
|
9050-0011 |
PREP-FIX flow cell for prep-HPLC, set at factory 0.1-3 mm, 10 uL, 1000 psi |
|
9050-0013 |
MICRO flow cell for micro-LC, 2 mm, 1.0 uL, 1000 psi |
|
9050-0009 |
STD-PREP flow cell for prep-LC or FIA, 10 mm, 20 uL, low back-pressure |
|
9050-0015 |
PREP flow cell for prep-LC or FIA, 5 mm, 10 uL, low back-pressure |
|
Accessories |
|
|
9100-0001 |
Deuterium Lamp Module for VUV-14D, VUV-24 |
|
9100-0002 |
Mercury Lamp Module for VUV-14 |
|
9100-0003 |
Tungsten Lamp Module for VUV-14 |
|
9150-0xxx |
Interference Filter 214-365 Nm for VUV-14, xxx denotes wavelength |
|
9150-0xxx |
Interference Filter 366-940 Nm for VUV-14, xxx denotes wavelength |
|
9110-0003 |
RJ-45 to DB-9 adaptor, female RJ-45 to Female DB-9, for serial connections | | |