분 주 량  : 0.5 uL ~ 5.0 mL
 분주시간 : 8.3 uL/sec ~ 2.5 mL/sec
 분주간격 : 0.5~999 sec
 분주회수 : 0~30,000 cycle

 

  • HMD0는 정밀주사기펌프를 이용하여 제작된 정밀 분주기, 충전시스템입니다.
  • 소량의 시료를 매우 정밀하게 분주하거나 충전할 수 있도록 설계되었다.
  • HMD04-1 모듈는 4-채널 로 구성되며, 4개의 유닛을 구성하면 16-채널까지 확장할 수 있다.
  • The execution unit of the filling system uses high-precision syringe pump industry, more suitable for filling small doses precisely
  • Control the multi-channel
  • Can be set cycles, the interval time, channels can be corrected individually, the convenience of customers
  • Select internal/external control mode through the button
  • Receiver filling start signal, filling function
  • Receives signals lack irrigation bottle stopper realize lack of a bottle filling features
  • Large-screen LCD, friendly interface
  • Pretreatment reflux, flushing function
  • Can memorize a variety of work settings parameter
  • Housing has certain corrosion, liquid splash function, protection class IP31

 

HMD04-1


Dispensing volume

100 uL ~ 5 mL (20-100% of syringe capacity)

Accuracy

≤1%(100% of syringe capacity)

Precision

≤0.5%(100% of syringe capacity)
≤2%(20% of syringe capacity)

Filling speed

8.3 uL/sec ~ 2.5 mL/sec

Filling time

Automatically calulate the time based on volume and speed

Pause time

0-3,600 sec

Copy number

0-30,000 times, "0" means unlimited cycle

Calibration

Each cjannel dispensing colume is calibrated separately

Rince time

1-100 times

Syringe

500 μL ~ 5 mL, full stroke is 30mm

Wetted material

Borosilicate glass, PTFE, PEEK

Pressure

Max. 0.68 MPa

Syringe fitting

1/4"-28 thread

Tubing fitting

1/4"-28 thread

Control mode

Control panel or external control

Dispensing signal input

TTL signal, high level to trigger (all channels are controlled synchronously)

Rincing signal input

TTL signal, high level to trigger (all channels are controlled synchronously)

No bottle signal input

TTL signal, high level to trigger (all channels are controlled synchronously)

Dimension

600 mmL x 256 mmW x 170 mmH (controller and drive unit)

Weight

10 kg (controller and drive unit)

Power

AC 90-260V/60 W

IP rating

IP31, Stainless steel housing

Model

Code

Stroke

Channel

Voltage

Current

Diemension

Weight

HMD04-1 05.11.10A 30 mm 4 90-260 VAC 60 W 600x256x170 10 kg